The Dwyer Series 2-5000 MiniHelic® II is a compact, dial-type differential pressure gage designed for clean, cost-effective measurement in applications requiring no external power. The series supports room positive pressure sensing, cabinet air-purging, medical respiratory equipment, air samplers, electronic air cooling systems, laminar flow hoods, filter status indication, fume hood face velocity monitoring, and duct pressures.
| Accuracy | ± 5 % |
|---|---|
| Accuracy Detail | ± 5% of FS at 70°F (21.1°C). |
| Compliance Note | Meets the technical requirements of EU Directive 2011/65/EU (RoHS II). |
| Dial Dia | 2 1/16 in |
| Dual | false |
| Materials | Glass Filled Polyamide, Polycarbonate |
| Mounting Orientation | Vertical Diaphragm |
| Operating Temperature, Max | 120 °F |
| Operating Temperature, Min | 20 °F |
| Pressure Limits | Up to 30 psi |
| Measurement Pressure, Min | 0 inH₂O |
| Measurement Pressure Type | Differential |
| Process Connection Note | Barbed, for 3/16˝ ID tubing (standard); 1/8˝ male NPT (optional). |
| Series ID | Series-2-5000 |
| Weight | 170.1 g |
| Certification | The Dwyer Series 2-5000 MiniHelic® II is a compact, dial-type differential pressure gage designed for clean, cost-effective measurement in applications requiring no external power. The series supports room positive pressure sensing, cabinet air-purging, medical respiratory equipment, air samplers, e |
| Process connection | Service is restricted to air and compatible non-corrosive gases. The gage features a blow-out membrane molded in conjunction with the diaphragm to accommodate over-pressure events safely up to 30 psig (2.067 bar) continuous total pressure on either connection. |
| Calibration | The gage is calibrated and zeroed at the factory in a vertical position. If used in any other orientation, it must be re-zeroed each time the position changes; models with ranges under 5 inches w.c (1.24 kPa) should generally remain in the vertical position unless special calibration was specified. |
| Measuring range | The Series 2-5000 is available in multiple model numbers corresponding to specific measurement ranges, such as 2-5001 (0 to 1 in w.c), 2-5005 (0 to 5 in w.c), and metric variants like 2-5000-1KPA. Models are offered with standard barbed connections or the -NPT suffix for threaded fittings. |
| Dimensions | Dimensions vary slightly by configuration but generally feature a diameter of approximately 2-19/32" (65.88 mm) and an overall height including mounting flange of 3-1/16" (77.79 mm). The unit weighs 6 oz (170 g). |
| Units available | bar, °f, °c, kpa |
Send the model code or describe the application — we reply within one working day.